Problem: No Display Common Cause: No power to the meter
Incorrect power to the meter
Damaged meter Check:
Voltage (90 to 264 VAC) across rear connector terminal 11 & 12. Return to Factory.
Problem: Incorrect Setpoint Common Cause:
Alarm not set correctly Check:
Alarm type, NO or NC & Hysteresis and setpoint setting.
Problem: Relay Output Inoperative Common Cause:
Incorrect connections
Contacts shorted or open due to electrical overload Check: Wiring Diagram Return to Factorys
Problem: Relay Chatter Common Cause: Electrical pickup Check:
See "SETPOINT RELAY/ALARMS" & "SIGNAL NOISE ISOLATION"
Problem: Analog Output Does Not Track Display Common Cause:
Incorrect setup
Defective Analog P.C.B. Check:
See flow diagram & "CALIBRATION" Replace Analog Printed Circuit Board
Problem: Display Drifting Common Cause:
Water in cable Check:
With transmitter wires disconnected, resistance between the red or black lead and the shield should be more than 15 Meg ohms